A Revolutionary Sensor
The new CHEMTROL PPM Sensor is a solid state sensor with a selective membrane that provides direct readings of Free Chlorine in water. CHEMTROL is introducing new PPM sensors for direct measurement of FREE CHLORINE in water in parts per million (ppm) or milligrams per liter (mg/l). Unlike other chlorine sensors on the market, these sensors are very precise and so reliable. They are easy to install and practically maintenance free. For best water management, it is recommended to use ORP control of chlorine oxidizing activity and PPM monitoring of free chlorine. Alternatively, PPM control can be used for control of the sanitizer and ORP control for a non-chlorine oxidize.
Standard Features
- Direct FREE CHLORINE measurement including: - free chlorine (HOCl and OCL-) - combined isocyanurates if present, corresponding to the DPD1 test pH RANGE of 5.5 to 9.5
- Temperature range of 5 to 50C (40 to 120 F)
- Cyanuric acid compatible
- Pressure range up to 3 bars (45 psi)
- PVC body with 1" MPT fitting,
- Replaceable membrane and electrolyte
- Flow cell with specially designed probe housing
- Easy calibration on startup
The chlorine sensor is shipped in a molded foam package containing:
- One PPM sensor with serial number (needed for warranty)
- One compression fitting with O-ring and compression disk
- One electrolyte cap with membrane
- One electrolyte bottle and screw-on filling tip
- One screwdriver (not needed if pre-wired)
A Choice of Ranges
The following standard ranges are available - P/N PPM002 from 0.02 to 2 ppm, - P/N PPM010 from 0.1 to 10 ppm. Other sensors are available for measurements up to 200 ppm.
PPM, ORP and pH
For best results in water management, it is recommended to use both ORP control of free chlorine and PPM monitoring of total and combined chlorine. Alternatively, PPM control can be used for control of the sanitizer and ORP control for an oxidizing agent, such as ozone.
Sensor Calibration
To calibrate the sensor, use a DPD or OTO test kit for total chlorine. It is recommended to do the first calibration 2 hours after startup and to repeat it after 24 h. After that, check the calibration as required and generally every 3 to 4 weeks.
Sensor Maintenance
There is no special maintenance requirements except to make sure that the membrane remains clean and to replace the membrane and electrolyte when needed. In winter, remove the sensor to prevent freezing if applicable. Do not reuse the membrane once removed. Replace with new membrane and electrolyte. In order to prevent serious damage to the sensor, be careful not to touch the membrane or the electrodes. To facilitate shipment and storage of the sensor, it is not fitted with the electrolyte cap until ready for installation. The sensor cap must be filled with the electrolyte solution before usage and emptied during storage.